This is the preview version of the Wisconsin State Legislature site.
Please see http://docs.legis.wisconsin.gov for the production version.
NR 149.35(1)(1)Scope. This subchapter establishes personnel, quality assurance, quality control, method selection, sample handling, and documentation requirements for laboratories.
NR 149.35(2)(2)Responsibility for quality system. A laboratory shall conduct analytical activities under a quality system that incorporates the provisions of this subchapter. At least one individual within a laboratory’s organization or under the laboratory’s employment shall be identified to the department as responsible for establishing, implementing, assessing, and revising, as needed, a laboratory’s quality system.
NR 149.35 HistoryHistory: CR 17-046: cr. Register February 2021 No. 782, eff. 6-29-21.
NR 149.36NR 149.36Laboratory personnel.
NR 149.36(1)(1)Management and analytical staff. The laboratory shall have personnel with education, training, or experience that allows the laboratory to comply with the requirements of this chapter. Contractors, external to the laboratory, may serve in key laboratory roles. When external contractors serve in essential laboratory roles, the contracts shall be available to the department to ensure that contractual specifications satisfy the requirements of this chapter.
NR 149.36 NoteNote: For requirements regarding changes in key personnel see s. NR 149.155.
NR 149.36(2)(2)Demonstration of capability.
NR 149.36(2)(a)(a) When a laboratory references a method that contains procedures for demonstrating initial capability, continuing capability or both, personnel performing analyses using these methods shall perform the procedures, meet any associated evaluation criteria, and document the results. When initial demonstrations of capability include the analysis of samples, the samples shall be prepared from a clean matrix and processed through all method preparation steps.
NR 149.36(2)(b)(b) When a laboratory references a method that does not contain procedures for demonstrating initial capability, the laboratory shall establish initial demonstration of capability criteria for determining that each person who performs testing on compliance samples using the method has demonstrated the necessary skills and expertise required to generate quality analytical results. The laboratory shall retain documentation that each person performing a given test on compliance samples has satisfied the demonstration of capability criteria established by the laboratory.
NR 149.36 HistoryHistory: CR 17-046: cr. Register February 2021 No. 782, eff. 6-29-21.
NR 149.365NR 149.365Laboratory ethics. All the following practices are prohibited and may result in enforcement action under s. NR 149.10:
NR 149.365(1)(1)Fabrication, falsification, or misrepresentation of data.
NR 149.365(2)(2)Improper instrument clock setting, termed time traveling, or improper recording of date or time.
NR 149.365(3)(3)Unwarranted manipulation of samples, software, peak integration, or analytical conditions.
NR 149.365(4)(4)Concealing or failing to report a known improper or unethical behavior or action associated with sample analysis.
NR 149.365 HistoryHistory: CR 17-046: cr. Register February 2021 No. 782, eff. 6-29-21.
NR 149.37NR 149.37Quality manual.
NR 149.37(1)(1)Purpose and general provisions. Each laboratory shall define its quality system in a quality manual. All policies and procedures governing the laboratory’s quality system shall be documented or referenced in the quality manual. All laboratory personnel shall follow the policies and procedures established by the quality manual.
NR 149.37(2)(2)Format. The quality manual shall have a format that addresses the content elements specified in this section. Content elements may be presented in narrative, tabular, schematic, or graphical form. The manual shall be a document in hard copy or electronic format traceable to the laboratory.
NR 149.37(3)(3)Content. Unless included in other standard operating procedures maintained under s. NR 149.40, the quality manual shall include, address, or refer to all the following elements:
NR 149.37(3)(a)(a) Procedures for retention, control, and maintenance of documents used in or associated with analysis.
NR 149.37(3)(b)(b) Procedures for achieving traceability of standards, reagents, and reference materials used to derive any results or measurements.
NR 149.37(3)(c)(c) Procedures for handling samples.
NR 149.37(3)(d)(d) Procedures for calibration, verification, and maintenance of support equipment.
NR 149.37(3)(e)(e) Procedures for evaluating quality control samples.
NR 149.37(3)(f)(f) Procedures for initiating, following up on, and documenting corrective action, addressing quality assurance and quality control failures, and any discrepancies or nonconformances.
NR 149.37(4)(4)Revisions. The quality manual shall be kept current. All editions or versions of the quality manual shall indicate the dates in which the quality manual was issued or revised.
NR 149.37 HistoryHistory: CR 17-046: cr. Register February 2021 No. 782, eff. 6-29-21.
NR 149.38NR 149.38Corrective action.
NR 149.38(1)(1)The laboratory shall take corrective action in response to any nonconformances including all the following:
NR 149.38(1)(a)(a) Departures from established procedures in the quality system are identified.
NR 149.38(1)(b)(b) Quality control samples fail, unless immediate reanalysis of the affected sample resolves the issue.
NR 149.38(2)(2)The corrective action under sub. (1) shall identify the problem, determine the most probable cause of the problem, implement solutions to correct the problem, and include a mechanism to verify that the action has had the desired effect.
NR 149.38(3)(3)The laboratory shall document corrective action taken to address the nonconformance under sub. (1) and any other changes resulting from corrective action investigations. Changes implemented to address failures of quality control samples shall be those that resolve or address the failure. Changes shall be implemented to minimize the number of affected results reported by a laboratory.
NR 149.38(4)(4)The laboratory shall monitor the effectiveness of implemented corrective action changes and take additional corrective action when initial or subsequent corrective action fails to resolve the nonconformance.
NR 149.38 NoteNote: The analyst may not always be able to identify the cause of isolated nonconformance incidents.
NR 149.38(5)(5)Root cause analysis shall be performed when there is recurrence.
NR 149.38 HistoryHistory: CR 17-046: cr. Register February 2021 No. 782, eff. 6-29-21.
NR 149.39NR 149.39Records and documents.
NR 149.39(1)(1)Records and documents retention and control.
NR 149.39(1)(a)(a) The laboratory shall establish procedures to control and manage all records and documents that form part of its quality system and that are required to demonstrate compliance with this chapter.
NR 149.39(1)(b)(b) The procedures shall ensure that documents required to perform analyses and to ensure the quality of generated data are available to laboratory personnel, and that records and documents are reviewed periodically for continuing suitability and, when necessary, revised to facilitate compliance with the requirements of this chapter.
NR 149.39(1)(c)(c) The laboratory shall retain all records and documents, which are part of its quality system, and that are required to demonstrate compliance with this chapter, for a minimum of three years after the generation of the last entry in an associated record or document. The laboratory shall retain records and documents for a longer minimum period if the records and documents are necessary to reconstruct analytical results generated during a three-year period.
NR 149.39(1)(d)(d) The department may require, in writing, that records be retained for a longer period than that specified in par. (c) if the department has initiated legal action involving test results or the accreditation status of the laboratory.
NR 149.39(1)(e)(e) The laboratory shall identify to the department a responsible party for retaining documents and records for the required period in the event the laboratory changes ownership or ceases to be accredited.
NR 149.39(1)(f)(f) Records and documents shall be handled and stored in a manner that ensures permanence and security for the required retention period and that facilitates retrieval to demonstrate compliance with this chapter.
NR 149.39(1)(g)(g) All records shall allow for reconstruction of reported results from raw data. Records and documents shall be legible, and entries shall be safeguarded against obliteration, erasures, overwriting, and corruption and are subject to all the following requirements:
NR 149.39 NoteNote: The determination of legibility includes concerns regarding the quality and permanence of records and the ability to decipher numbers and letters. For example, thermal paper ages and eventually becomes unreadable, so thermal paper printouts should ultimately be scanned or copied to ensure permanence.
NR 149.39(1)(g)1.1. Handwritten records shall be recorded in ink.
NR 149.39(1)(g)2.2. Records and documents that are stored only on electronic media shall be supported by the hardware and software necessary for retrieval and reproduction into hard copy.
NR 149.39(1)(g)3.3. Corrections or other alterations made to entries in records or documents may not obscure the original entry.
NR 149.39(1)(g)4.4. The laboratory shall have procedures to prevent unauthorized access or amendments to records and documents.
NR 149.39(2)(2)Administrative records. A laboratory shall maintain all the following administrative records:
NR 149.39(2)(a)(a) Certificates of accreditation issued by the department unless the department has requested a laboratory to return the certificates to the department.
NR 149.39(2)(b)(b) Certificates issued to the laboratory by entities with which the department has entered into a reciprocal agreement under s. NR 149.08, if a laboratory is accredited for this chapter under any existing agreement.
NR 149.39(2)(c)(c) Records of personnel qualifications, experience, and training when personnel are required to possess or maintain specific credentials by s. NR 149.36 (2).
NR 149.39(2)(d)(d) Copies of, or access to, other regulations, standards, and documents necessary for the laboratory to operate or to maintain compliance with this chapter.
NR 149.39(3)(3)Reagent and standard records and reference materials. The laboratory shall document the identity, source, and purity of standards and reagents used in the methods performed.  The laboratory shall retain records of reference materials and certificates of analysis when the records are provided by the supplier and are necessary to establish the identity, source, or purity of standards and reagents.
NR 149.39(3)(a)(a) Reagent containers shall be labeled with an expiration date, chemical name, and concentration. Except for instrument vials, standard containers shall be labeled with an expiration date, chemical name, and concentration.
NR 149.39(3)(b)(b) The laboratory shall document the lot number, manufacturer, chemical name, concentration, and the date of expiration for standards and reagents purchased from a manufacturer. These records shall be separate from the container labels.
NR 149.39 NoteNote: An expiration date is not required when one is not provided by the supplier.
NR 149.39(3)(c)(c) The laboratory shall document the preparation details of all prepared standards and reagents. These records shall link the prepared standards and reagents to the respective originating stocks or neat compounds and shall indicate the date of preparation, date of expiration, and the identity of the preparer.
NR 149.39(3)(d)(d) The laboratory may not use any standards and reagents beyond the expiration dates unless the laboratory is using the standard and reagents for qualitative determinations.
NR 149.39(3)(e)(e) Certificates for all reference materials shall be maintained.
NR 149.39(4)(4)Analytical and technical records. The format of the analytical and technical records of a laboratory shall facilitate access to the information in this subsection and may be contained in bench sheets, log books, notebooks, journals, manuals, standard operating procedures under s. NR 149.40, and forms, in hard copy or electronic media.
NR 149.39(5)(5)Sample collection records. The laboratory shall retain records supplied by the collector to allow the laboratory to evaluate collection information against the laboratory’s sample acceptance policy.
NR 149.39 HistoryHistory: CR 17-046: cr. Register February 2021 No. 782, eff. 6-29-21.
NR 149.40NR 149.40Standard operating procedures.
NR 149.40(1)(1)A laboratory shall maintain written standard operating procedures that document or reference activities needed to maintain its quality systems and that enable performing or reproducing an analysis in its entirety as performed at the laboratory. Each laboratory shall develop, maintain, and keep current its standard operating procedures for both sample preparation and analysis.
NR 149.40 NoteNote: Sample preparation includes digestions, distillations, extractions, concentrations, dilutions, and clean-up performed on samples prior to the determinative analytical step.
NR 149.40(2)(2)Standard operating procedures may be documents written by laboratory personnel or may consist entirely of copies of published documents, manuals, or procedures if the laboratory follows the chosen source exactly.
NR 149.40(3)(3)Standard operating procedures may consist, in part, of copies of published documents, manuals, or procedures if all the following conditions are met:
NR 149.40(3)(a)(a) Modifications to the published source are described in writing in additional documents.
NR 149.40(3)(b)(b) Clarifications, changes, or choices are completely described in additional documents, when published sources offer multiple options, ambiguous directives, or insufficient detail to perform or reproduce an analysis.
NR 149.40(4)(4)Standard operating procedures shall indicate the dates of issue or revision.
NR 149.40(5)(5)When the standard operating procedure is written by the laboratory, each standard operating procedure shall include, address, or refer to all the following elements, if applicable:
NR 149.40(5)(a)(a) Identification of the referenced method.
NR 149.40(5)(b)(b) For multi-analyte methods, a list of analytes.
NR 149.40(5)(c)(c) Potential interferences and how the interferences are treated.
NR 149.40(5)(d)(d) Equipment and analytical instruments.
NR 149.40(5)(e)(e) Consumable supplies, reagents, and standards.
NR 149.40(5)(f)(f) Sample preservation, storage, and hold time.
NR 149.40(5)(g)(g) Quality control samples and frequency of the analysis.
NR 149.40(5)(h)(h) Calibration and standardization.
NR 149.40(5)(i)(i) Procedure for analysis.
NR 149.40(5)(j)(j) Data assessment and acceptance criteria for quality control measures.
NR 149.40(5)(k)(k) Corrective actions and contingencies for handling out of control or unacceptable data.
NR 149.40 HistoryHistory: CR 17-046: cr. Register February 2021 No. 782, eff. 6-29-21.
NR 149.41NR 149.41Method selection.
Loading...
Loading...
Published under s. 35.93, Stats. Updated on the first day of each month. Entire code is always current. The Register date on each page is the date the chapter was last published.