This is the preview version of the Wisconsin State Legislature site.
Please see http://docs.legis.wisconsin.gov for the production version.
DHS 157.85(15)(a) (a) Periodic quality control checks shall be performed on all therapeutic radiation machines at intervals recommended by the manufacturer or by recognized national or international organizations.
DHS 157.85 Note Note: An acceptable reference is “Comprehensive QA for Radiation Oncology: Report of AAPM Radiation Therapy Committee Task Group 40," AAPM Report No. 46, American Association of Physicists in Medicine, April, 1994.
DHS 157.85(15)(b) (b) Quality control checks shall include determination of central axis radiation output and a representative sampling of periodic quality control checks according to recommendations of national or international organizations. Representative sampling shall include all referenced periodic quality control checks in an interval not to exceed 14 consecutive calendar months.
DHS 157.85 Note Note: An acceptable reference is “Comprehensive QA for Radiation Oncology: Report of AAPM Radiation Therapy Committee Task Group 40," AAPM Report No. 46, American Association of Physicists in Medicine, April, 1994. The publication may be consulted at the Department of Health Services, Radiation Protection Section, 1 West Wilson St, Room 150, Madison WI 53702-0007. AAPM reports may be obtained from Medical Physics Publishing, 4513 Vernon Blvd., Madison WI 53705-4964 or ordered from their website: www.medicalphysics.org.
DHS 157.85(16) (16) Operating procedures.
DHS 157.85(16)(a)(a) A therapeutic radiation machine may not be left unattended unless secured.
DHS 157.85(16)(b) (b) When a patient must be held in position for radiation therapy, mechanical supporting or restraining devices or other means recommended by a physician shall be used.
DHS 157.85(16)(c) (c) An x-ray tube housing assembly may not be held by a person during operation unless the assembly is designed to require holding and the peak tube potential of the system does not exceed 50 kV. In these cases, the holder shall wear protective gloves and apron of not less than 0.5 millimeters lead equivalency at 100 kV.
DHS 157.85(16)(d) (d) A copy of the current operating and emergency procedures shall be maintained at the therapeutic radiation machine control console.
DHS 157.85(16)(e) (e) No person other than the patient may be in the treatment room during exposures from therapeutic radiation machines operating above 150 kV. At energies less than or equal to 150 kV, any person, other than the patient, in the treatment room shall be protected by a barrier sufficient to meet the requirements of s. DHS 157.22 (1).
DHS 157.85(16)(f) (f) A registrant shall promptly repair any system that is not operating properly.
DHS 157.85(16)(g) (g) A registrant for electronic brachytherapy shall do all of the following:
DHS 157.85(16)(g)1. 1. Ensure the electronic brachytherapy unit is inoperable, either by hardware or password, when unattended by qualified staff or service personnel.
DHS 157.85(16)(g)2. 2. Secure the unit, console, console keys and the treatment room when unattended or not in use.
DHS 157.85(16)(g)3. 3. Prevent dual operation of more than one radiation producing device in a treatment room, if applicable.
DHS 157.85(16)(g)4. 4. Create a written procedure for safe operation of each device.
DHS 157.85(16)(g)5. 5. Develop, implement and maintain written procedures for responding to an abnormal situation. The procedure shall include all the following:
DHS 157.85(16)(g)5.a. a. Instructions for responding to equipment failures and the names of the persons responsible for implementing corrective actions.
DHS 157.85(16)(g)5.b. b. The names and telephone numbers of the licensed practitioner, the medical physicist, the radiation safety officer and the manufacturer to be contacted if the unit or console operates abnormally.
DHS 157.85(16)(g)6. 6. Maintain a copy of the procedures required by subds. 4. and 5. at the unit console.
DHS 157.85(16)(g)7. 7. Ensure all of the following are done during treatment:
DHS 157.85(16)(g)7.a. a. Only individuals approved by the authorized user, radiation safety officer or medical physicist may be present in the treatment room.
DHS 157.85(16)(g)7.b. b. Protective shielding shall be available for persons in the treatment room.
DHS 157.85(16)(g)7.c. c. A radiation survey is performed when the unit and/or shielding is portable to verify proper shielding placement immediately upon initiation of treatment.
DHS 157.85(16)(g)7.d. d. A medical physicist, oncologist, or operator shall be present and at the controls throughout the course of the patient's treatment. A medical physicist and oncologist shall remain available during treatment.
DHS 157.85(16)(g)7.e. e. A medical physicist or operator shall monitor the position of all persons in the treatment room to prevent unshielded exposure.
DHS 157.85(16)(g)7.f. f. A medical physicist or operator shall monitor all entrances to prevent entering individuals from unshielded exposure.
DHS 157.85(16)(g)7.g. g. Only mechanical supporting or restraining devices may be used to hold a patient in position, when applicable.
DHS 157.85 History History: CR 01-108: cr. Register July 2002 No. 559, eff. 8-1-02; CR 09-062: cr. (13) (em), (14) (fm), (g) 6., (gm) and (16) (g), am. (14) (e) and (g) 4. Register April 2010 No. 652, eff. 5-1-10; CR 16-078: am. (3) (title), (13) (em) 2. to 7., (14) (gm) (intro.), cr. (14) (gr), am. (16) (g) 7. d. Register January 2018 No. 745, eff. 2-1-18.
DHS 157.86 DHS 157.86Registration of radiation machine facilities.
DHS 157.86(1)(1)Registration fees.
DHS 157.86(1)(a)(a) An annual registration fee shall be levied for each site registration under this section, according to the following schedule:
DHS 157.86(1)(a)1. 1. For a site having an ionizing radiation installation serving physicians and clinics, osteopaths and clinics, chiropractors or hospitals, the fee shall be $50 for each site and $50 for each x-ray tube.
DHS 157.86(1)(a)2. 2. For a podiatric or veterinary site having an ionizing radiation installation, the fee shall be $50 for each site and $50 for each x-ray tube.
DHS 157.86(1)(a)3. 3. For a dental site having an ionizing radiation installation, the fee shall be $50 for each site and $35 for each x-ray tube.
DHS 157.86(1)(a)4. 4. For an industrial, school, research project or other site having an ionizing radiation installation, the fee shall be $50 for each site and $50 for each x-ray tube.
DHS 157.86(1)(a)5. 5. An additional fee of $50, regardless of the number of devices, shall be required for each registration whenever the annual fee for renewal is not paid prior to the expiration of the registration.
DHS 157.86(1)(a)6. 6. A change of ownership requires re-registration and fees paid by the new registrant.
DHS 157.86(1)(a)7. 7. Any change in registration information shall be submitted to the department within 30 days after the change takes place. No fee is required for recording changes in registration information.
DHS 157.86(1)(a)8. 8. Manufacturing, testing or servicing facilities shall be considered as one x-ray tube for registration purposes.
DHS 157.86(1)(a)9. 9. Electron microscopes and extremity bone densitometers are exempt from registration fees after the initial registration.
DHS 157.86(2) (2) Exemptions. The following items are exempted from the requirements of this section:
DHS 157.86(2)(a) (a) Electronic equipment that produces radiation incidental to its operation for other purposes, such as x-rays from radio or television transmitter high voltage tubes.
DHS 157.86(2)(b) (b) Radiation machines in transit or storage.
DHS 157.86(2)(c) (c) Domestic television receivers and computer monitors.
DHS 157.86(3) (3) Reciprocal recognition of out-of-state radiation machines.
DHS 157.86(3)(a)(a) Whenever any radiation machine is to be brought into the state, for any temporary use, the person proposing to bring the machine into the state shall give written notice to the department by mail or facsimile at least 48 hours before the machine is to be used in the state. The notice shall include all the following information:
DHS 157.86(3)(a)1. 1. The type of radiation machine.
DHS 157.86(3)(a)2. 2. The nature, duration and scope of intended use.
DHS 157.86(3)(a)3. 3. The exact location or locations where the radiation machine is intended to be used.
DHS 157.86(3)(a)4. 4. States in which the machine is registered.
DHS 157.86(3)(b) (b) If, for a specific case, the 48-hour notice period would impose an undue hardship on the person, that person may apply to the department for verbal permission to proceed sooner.
DHS 157.86 Note Note: The department may be contacted by phone at 608-267-4784 or facsimile at 608- 267-4799.
DHS 157.86(3)(c) (c) The person in control shall do all the following:
DHS 157.86(3)(c)1. 1. Comply with all applicable rules of the department.
DHS 157.86(3)(c)2. 2. Supply the department with other information as the department requests.
DHS 157.86(3)(c)3. 3. Not operate within the state on a temporary basis in excess of 30 calendar days per year without obtaining a Wisconsin registration.
DHS 157.86 History History: CR 01-108: cr. Register July 2002 No. 559, eff. 8-1-02; CR 06-021: am. (1) (a) 1. to 5. Register October 2006 No. 610, eff. 11-1-06; correction in (1) (a) 2. made under s. 35.17, Stats., Register January 2018 No. 745.
subch. IX of ch. DHS 157 Subchapter IX — Radiation Generating Devices
DHS 157.87 DHS 157.87Radiation safety requirements.
DHS 157.87(1)(1)General requirements. Unless otherwise provided in this chapter, this subchapter applies to all radiation generating devices. Certified and certifiable cabinet x-ray systems shall also meet the requirements of 21 CFR 1020.40.
DHS 157.87(1)(a) (a) Warning devices.
DHS 157.87(1)(a)1.1. Warning devices shall be labeled so that their purpose is easily identified.
DHS 157.87(1)(a)2. 2. An easily visible warning device light labeled with the words “X-RAY ON," or words having a similar meaning and intent, shall be located near any switch that energizes an x-ray tube and shall be illuminated only when the tube is energized. This warning light shall be of a fail-safe design.
DHS 157.87(1)(b) (b) Labeling.
DHS 157.87(1)(b)1.1. All radiation generating devices equipment shall be labeled near any switch that energizes an x-ray tube with either:
DHS 157.87(1)(b)1.a. a. A readily visible and discernible sign or signs bearing the radiation symbol
DHS 157.87(1)(b)1.b. b. A readily visible and discernible sign or signs bearing “CAUTION RADIATION - THIS EQUIPMENT PRODUCES RADIATION WHEN ENERGIZED," or words having a similar meaning and intent.
DHS 157.87(1)(b)2. 2. For radiation generating devices with designed openings for object entries, such as baggage security screening units, the following shall be posted at or near each opening: “CAUTION – X-RAY HAZARD: DO NOT INSERT ANY PART OF THE BODY WHEN SYSTEM IS ENERGIZED”, or words having similar meaning and intent.
DHS 157.87(1)(c) (c) Radiation source housing. Each x-ray tube housing shall comply with all of the following requirements:
DHS 157.87(1)(c)1. 1. `Interlock.' When the x-ray tube housing is the primary shielding for the x-ray tube, and is intended to be opened for normal use or maintenance, the housing shall be equipped with an interlock that shuts off the high voltage to the x-ray tube if the housing is opened.
DHS 157.87(1)(c)2. 2. `Radiation emission limit.' Each x-ray tube housing shall be so constructed that, with all shutters closed, the leakage radiation measured at a distance of 5 centimeters from the x-ray tube housing surface does not exceed 0.025 mSv (2.5 mrem) per hour. This limit shall be met at the maximum tube rating. For closed-beam systems, this requirement may be met by complying with the radiation emission limit provided in sub. (2) (d). For a radiation generating device in a shielded room, this limit may be met by measuring from any accessible surface outside the room housing the radiation generating device. For hand-held, open-beam radiation generating devices, this requirement may be met by complying with the limits in sub. (4) (c) radiation emission limit.
DHS 157.87(1)(d) (d) Generator cabinet or high voltage source radiation emission limits. Each x-ray generator or high-voltage source shall be supplied with a protective cabinet which limits leakage radiation to 2.5 microSv (0.25 mrem) per hour at a distance of 5 centimeters measured at the nearest accessible surface. For closed-beam systems, this requirement may be met by complying with radiation emission limits in sub. (2) (d). For a radiation generating device in a shielded room with the high-voltage generator also inside the shielded room, this limit may be met by measuring from any accessible surface outside the room housing the radiation generating device. For hand-held, open-beam radiation generating devices, this requirement may be met by complying with the radiation emission limits in sub. (4) (c).
DHS 157.87(1)(e) (e) Surveys.
DHS 157.87(1)(e)1.1. Radiation surveys of all radiation generating devices shall be sufficient to show compliance with radiation emission requirements of this subchapter and as required by ss. DHS 157.22 and 157.25. The radiation surveys shall be sufficient to evaluate the magnitude and extent of radiation emissions and the potential radiological hazards that could be present. At a minimum, surveys shall be performed:
DHS 157.87(1)(e)1.a. a. Upon installation of the equipment, and at least once every 12 months thereafter.
DHS 157.87(1)(e)1.b. b. Following any change in the initial arrangement, number, or type of local components in the system.
DHS 157.87(1)(e)1.c. c. Following any maintenance requiring the disassembly, removal, or repair of a local component in the system.
DHS 157.87(1)(e)1.d. d. During the performance of maintenance, calibration, and other procedures if the procedures require the presence of a primary x-ray beam while any local component in the system is disassembled or removed.
DHS 157.87(1)(e)1.e. e. After bypassing a safety device or interlock as required in par. (h).
DHS 157.87(1)(e)1.f. f. Any time a visual inspection of the local components in the system reveals an abnormal condition.
DHS 157.87(1)(e)1.g. g. Whenever a personnel monitoring device shows a significant increase over previous monitoring period or readings are approaching the limits specified in s. DHS 157.22.
DHS 157.87(1)(e)2. 2. The registrant shall have access to sufficiently calibrated, appropriate, and operable radiation survey instruments to make physical radiation surveys as required by this subchapter. The instruments shall be capable of detecting and measuring the types and levels of radiation involved including primary, scattered, and leakage radiation.
DHS 157.87(1)(e)3. 3. The registrant shall assure the maintenance and calibration of all monitoring and survey instruments used by the registrant or service provider per s. DHS 157.25.
DHS 157.87(1)(e)4. 4. The department may approve alternatives to the survey requirements of subd. 1., if the registrant submits to the department a request for approval of an alternative method that otherwise demonstrates compliance with this subchapter and ss. DHS 157.22 and 157.25.
DHS 157.87(1)(f) (f) Posting. Each area or room containing a radiation generating device where an individual may receive 0.02 mSv (2 mrem) in any one hour or 1 mSv (100 mrem) per year shall be conspicuously posted with a sign or signs bearing the radiation symbol described in s. DHS 157.29 (1) and the words "CAUTION - X-RAY EQUIPMENT," “CAUTION – RADIATION GENERATING DEVICE,” or words having a similar meaning and intent.
DHS 157.87(1)(g) (g) Security. Radiation generating devices shall be secured in such a way as to be only accessible to, or operable by, authorized personnel when not in operation.
DHS 157.87(1)(h) (h) Operating requirements.
DHS 157.87(1)(h)1.1. `Procedures.' Normal operating procedures shall be written and available to all radiation generating device workers. No individual shall be permitted to operate a radiation generating device in any manner other than that specified in the procedures unless the individual has obtained written approval of the person in control.
Loading...
Loading...
Published under s. 35.93, Stats. Updated on the first day of each month. Entire code is always current. The Register date on each page is the date the chapter was last published.