This is the preview version of the Wisconsin State Legislature site.
Please see http://docs.legis.wisconsin.gov for the production version.
DHS 157.87(1)(h)2. 2. `Bypassing.'
DHS 157.87(1)(h)2.a.a. No individual shall bypass a safety device or interlock, or remove shielding unless the individual has obtained the approval of the person in control. The approval shall be for a specified period of time.
DHS 157.87(1)(h)2.b. b. When a safety device or interlock has been bypassed, a readily discernible sign bearing the words "SAFETY DEVICE NOT WORKING," or words having a similar meaning and intent, shall be placed on the radiation source housing and at the control switch.
DHS 157.87(1)(h)2.c. c. A record of any bypass of a safety device or interlock shall be maintained. The record shall contain information such as the date the alteration was made, type of alteration, length of time the unit remained in the altered condition, and post bypass survey. The record shall be signed by the person in control, the individual who made the alteration, and the individual who restored the unit to original manufacturer's specifications.
DHS 157.87(1)(h)3. 3. `Control panel.'
DHS 157.87(1)(h)3.a.a. The radiation generating devices hall only be activated from a control panel.
DHS 157.87(1)(h)3.b. b. All indicators and controls that control the primary beam shall be identifiable and discernible through the use of labels, symbols, software displays or the equivalent.
DHS 157.87(1)(h)4. 4. `Interlocks.'
DHS 157.87(1)(h)4.a.a. An interlock shall not be used to de-activate the x-ray tube or radiation generating device, except in an emergency or during testing of the interlock system.
DHS 157.87(1)(h)4.b. b. After triggering any interlock, it shall only be possible to reset the radiation generating device to full operation from a control panel. All interlocks shall be of a fail-safe design.
DHS 157.87(1)(h)5. 5. `Multiple sources.' If more than one x-ray tube assembly or focal spot can be operated sequentially or simultaneously from a control panel, visual indicators shall identify which tube assembly or focal spot has been selected. The selectors shall be identified as to their function. If a letter or number is used, a reference card or table explaining the code shall be affixed to the control panel.
DHS 157.87(1)(i) (i) Repair or modification of x-ray tube or radiation generating device systems. Only trained personnel or registered service providers shall be permitted to install, repair, or make modifications to the radiation generating device. No operation involving removal of covers, shielding materials or tube housings or modifications to shutters, collimators, or beam stops shall be performed without ensuring that the tube is off and will remain off until safe conditions have been restored. The main power switch with a lockout tagout, rather than interlocks, shall be used for routine shutdown in preparation for repairs. It is the responsibility of the registrant to assure that qualified personnel install, repair, or make modifications to the radiation generating device.
DHS 157.87(1)(j) (j) Testing of safety devices.
DHS 157.87(1)(j)1.1. Tests of all safety devices, such as interlocks, shutters, warning lights, and required emergency shut-off switches shall be conducted at intervals not to exceed 6 months on all operable radiation generating devices.
DHS 157.87(1)(j)2. 2. If any safety device fails during testing, the radiation generating device shall be removed from service until the safety device failure is corrected or proper temporary administrative controls are established and approved in writing by the person in control.
DHS 157.87(1)(j)3. 3. Records of safety device tests, check dates, findings and corrective actions shall be available for inspection and maintained for 5 years.
DHS 157.87(1)(j)4. 4. Records shall include the date of each test, a list of the safety devices tested, survey instrument information, calibration date, the results of each test, the name of the person performing each test and corrective actions taken for safety devices that fail each required test.
DHS 157.87(1)(j)5. 5. Testing of safety devices may be deferred if the unit or installation is clearly marked and kept out of service. Units or installations brought back into service after exceeding a 6 month interval shall be tested prior to use.
DHS 157.87(1)(j)6. 6. If testing of a safety device cannot be performed due to manufacturer design, the registrant shall document that the safety device will not be tested and why the safety device cannot be tested.
DHS 157.87(1)(k) (k) Instruction and training. The registrant shall document the scope of training required for the radiation generating device they possess in accordance with this section. No individual shall be permitted to operate or maintain a radiation generating device, or enter a shielded room without appropriate instruction and training. Records shall be maintained onsite of all required training and instruction, and made available for review by the department. Each individual permitted to operate or maintain a radiation generating device shall receive instruction, hands-on training with the radiation generating device and equipment, and demonstrate competence in all of the following:
DHS 157.87(1)(k)1. 1. Types of radiation and identification of radiation hazards associated with the use of the radiation generating device and associated equipment and precautions or measures to take to minimize radiation exposure.
DHS 157.87(1)(k)2. 2. Significance of the various radiation warning, safety devices, and interlocks incorporated into the equipment, or the reasons they have not been installed on certain pieces of equipment and the extra precautions required in those cases.
DHS 157.87(1)(k)3. 3. Commensurate with potential hazards of use, biological effects of radiation, radiation risks, and recognition of symptoms of an acute localized exposure.
DHS 157.87(1)(k)4. 4. Normal operating procedures and procedures to prevent unauthorized use for each type of radiation generating device and associated equipment.
DHS 157.87(1)(k)5. 5. Procedures for reporting an actual or suspected accidental exposure or other radiation safety concerns, such as any unusual occurrence or malfunction that may involve exposure to radiation.
DHS 157.87(1)(k)6. 6. Performing surveys where applicable.
DHS 157.87(1)(L) (L) Radiation protection responsibility.
DHS 157.87(1)(L)1.1. The registrant's management shall make the final decision to use any radiation generating device and bear responsibility for radiation safety.
DHS 157.87(1)(L)2. 2. The registrant's management shall designate an individual responsible for radiation safety. This individual shall have direct access to management for radiation safety issues. This individual shall have training and experience commensurate with the scope of the radiation safety program to carry out all of the following responsibilities:
DHS 157.87(1)(L)2.a. a. Ensuring that all radiation generating devices are operated within the limitations of the established radiation safety program and operating procedures.
DHS 157.87(1)(L)2.b. b. Instructing personnel with regard to safe working practices and ensuring all personnel are trained in radiation safety commensurate with the hazards of the job.
DHS 157.87(1)(L)2.c. c. Investigating any incident of abnormal operation or exposure or suspected overexposure of personnel to determine the cause, take remedial action, and report the incident to the proper authorities.
DHS 157.87(1)(L)2.d. d. Ensuring that safety devices, interlocks, warning signals, labels, postings, and signs are functioning and located where required.
DHS 157.87(1)(L)2.e. e. Maintaining all radiation safety records.
DHS 157.87(2) (2)Additional requirements for closed-beam radiation generating devices. In addition to the requirements of sub. (1), all of the following apply to all closed-beam x-ray radiation generating devices:
DHS 157.87(2)(a) (a) System enclosure. The radiation source, sample or object, detector, and, if used, analyzing crystal shall be enclosed in a chamber or coupled chambers that cannot be entered by any part of the body during normal operation.
DHS 157.87(2)(b) (b) Interlocks. All doors and panels accessing the radiation generating device shall be interlocked. The interlocks shall be of a fail-safe design.
DHS 157.87(2)(c) (c) Interlock functions. The system enclosure and sample chamber closure shall be interlocked with the x–ray tube high voltage supply or a shutter in the primary beam so that no x-ray beam can enter the sample or object chamber while it is open unless the interlock has been conspicuously and deliberately defeated. The interlock shall be of fail-safe design or adequate administrative controls shall be exercised to ensure operations will not continue without a proper functioning interlock.
DHS 157.87(2)(d) (d) Radiation emission limit. The radiation emission for all closed beam radiation generating devices shall not exceed a dose rate of 0.005 mSv (0.5 mrem) in one hour at five centimeters outside any accessible surface.
DHS 157.87(2)(e) (e) Security screening units. Security screening units shall be provided with means to ensure operator presence at the control area in a position which permits surveillance of the openings and doors during generation of x-ray radiation. The means provided shall meet any of the following conditions:
DHS 157.87(2)(e)1. 1. During an exposure or preset succession of exposures of one-half second or greater duration, the operator is able to terminate the exposure or preset succession of exposures at any time.
DHS 157.87(2)(e)2. 2. During an exposure or preset succession of exposures of less than one-half second duration, the operator is able to prevent additional exposures after the completion of the exposure in progress.
DHS 157.87(3) (3)Additional requirements for open-beam radiation generating devices. The registrant shall submit a written explanation of the need to utilize the open beam configuration instead of closed-beam systems and receive the written approval of the department prior to use. In addition to the requirements in sub. (1) all of the following apply to all open-beam radiation generating devices not otherwise addressed in this subchapter:
DHS 157.87(3)(a) (a) Safety device.
DHS 157.87(3)(a)1.1. The registrant shall use a safety device which prevents the entry of any portion of the operator's body into the path of the primary beam or which causes the primary beam to be shut off upon entry into its path, unless otherwise authorized by the department or this subchapter.
DHS 157.87(3)(a)2. 2. If the registrant's use of the open-beam radiation generating device does not permit the use of a safety device described in subd. 1. or 4., the registrant shall submit a written justification for why such a safety device cannot be used and receive the written approval of the department prior to use. The documented request and approval, include information about the absence of safety devices, shall be available for inspection as long the open-beam radiation generating device is in use. A record of this documentation shall be retained for 5 years after the open-beam radiation generating device is used.
DHS 157.87(3)(a)3. 3. If the registrant does not use a safety device described in subd. 1., the registrant shall employ alternative methods such as policies and procedures to minimize the possibility of unnecessary exposure. These alternative methods shall be documented and available for inspection as long as these methods are employed. A record of this documentation shall be retained for 5 years after the method is no longer employed.
DHS 157.87(3)(a)4. 4. For portable open-beam radiation generating devices that are manufactured to be used hand-held, or potentially used as a hand-held, without a safety device described in subd. 1, this safety device requirement shall be met by complying with all the requirements in sub. (4) prior to use.
DHS 157.87(3)(b) (b) X-ray on status. For open-beam equipment, radiation generating devices shall be provided with a readily discernible and active indication of all of the following:
DHS 157.87(3)(b)1. 1. `X-ray tube “on-off" status.' An on-off status indicator shall be of a fail-safe design and be located near the radiation source housing. The warning lights as required by sub. (1) (a) 2. may meet this requirement if the warning lights are readily discernible and viewable by anyone near the primary beam.
DHS 157.87(3)(b)2. 2. `Shutter “open-closed" status.' An open-closed status indicator shall be of a fail-safe design and be located at the control panel. If the primary beam is controlled with a shutter, it shall be near each beam port on the radiation source housing. The shutter status indicator shall be clearly labeled as to the meaning of the status of the device. The status indicator at the control panel may meet the requirement for the status indicator at the beam port if the status indicator at the control panel is readily discernible and viewable by anyone near the primary beam.
DHS 157.87(3)(c) (c) Labeling. Each unit shall be labeled at or near the x-ray exit beam port to identify the location of the beam with the words, “CAUTION - X-RAY BEAM", “CAUTION – HIGH INTENSITY X-RAY BEAM”, or words having a similar meaning and intent.
DHS 157.87(3)(d) (d) Beam ports. Unused beam ports on radiation source housings shall be secured in the closed position in a manner which will prevent inadvertent opening.
DHS 157.87(3)(e) (e) Shutters. On open-beam radiation generating device configurations that are designed to accommodate interchangeable components, such as apertures, filters, or target materials, each beam port on the radiation source housing shall be equipped with a shutter that cannot be opened unless a collimator or a component coupling has been connected to the beam port.
DHS 157.87(3)(f) (f) Radiation emission limits. The local components of an open-beam radiation generating device, such as portable shielding, room walls, and designed housing, shall be located and arranged and shall include sufficient shielding or access control such that no radiation emissions exist, exclusive of the primary beam, in any area surrounding the local component group which could result in a dose to an individual present therein in excess of the dose limits as outlined in s. DHS 157.25. These emissions shall be met at any specified tube rating.
DHS 157.87(3)(g) (g) Primary beam attenuation. In cases where the primary x-ray beam is not intercepted by the detector device under all conditions of operation, protective measures shall be provided, such as auxiliary shielding or administrative procedures, to avoid exposure to any individual from the transmitted primary x-ray beam.
DHS 157.87(3)(h) (h) Operator attendance. The operator shall be in immediate attendance at all times when the equipment is in operation except when the area is locked, or the equipment is secured to protect against unauthorized or accidental entry.
DHS 157.87(3)(i) (i) Control of access. If the radiation generating device is not in a restricted area, the operator shall be able to control access to the radiation generating device at all times during operation. If the radiation generating device is not in a restricted area and the radiation generating device is capable of creating a radiation area or, the operator shall be able to control access to the radiation generating device at all times during operation, and do all the following:
DHS 157.87(3)(i)1. 1. Conspicuously identify radiation areas. The radiation source shall be within a conspicuous perimeter that identifies the area in which the dose equivalent rate exceeds 0.05 mSv (5 mrem) per hour. The area identified shall be suitably posted with “CAUTION -RADIATION AREA" signs at the perimeter. The operator shall ensure that no one is inside or enters the radiation area during operation of the radiation generating device.
DHS 157.87(3)(i)2. 2. Conspicuously identify high radiation areas. The radiation source shall be within a conspicuous perimeter that identifies the area in which the dose equivalent rate exceeds 1 mSv (100 mrem) per hour. The area identified shall be suitably posted with “CAUTION - HIGH RADIATION AREA" signs at the perimeter. The operator shall ensure that no one is inside or enters the high radiation area during operation of the radiation generating device.
DHS 157.87(3)(i)3. 3. Perform a visual check of the controlled area to ensure it is free of all unauthorized personnel immediately prior to activating or exposing the radiation source.
DHS 157.87(3)(i)4. 4. Maintain surveillance of the exposure area during operation, either visually or by other reliable means to ensure that no person enters the area.
DHS 157.87(3)(i)5. 5. With the exception of hand-held x-ray systems, when approaching the radiation source, following the conclusion of an exposure, use a suitable calibrated and operable radiation detection instrument to verify that the radiation source is in its fully shielded condition or that the x-ray tube has been de-energized.
DHS 157.87(3)(i)6. 6. Wear a personal alarming dose rate meter to approach the work area. The radiation field in the work area must be uniform such that the personal alarming dose rate meter responds to radiation exposure to any part of the body. The personal alarming dose rate meter shall not be used to measure radiation levels required under subds. 7. to 9., nor shall it be used to indicate the presence of a radiation source for potential non-uniform exposure, such as may occur during machine maintenance or work in a radiation generating device target area. The personal alarming dose rate meter shall meet all of the following:
DHS 157.87(3)(i)6.a. a. Is appropriately designed and calibrated for the type of x-ray emitted, such as pulse or continuous.
DHS 157.87(3)(i)6.b. b. Is set at an appropriate level to detect the presence of the radiation source, for example 0.02 mSv (2 mrem) per hour.
DHS 157.87(3)(i)6.c. c. Has been source-checked for response prior to use.
DHS 157.87(3)(i)7. 7. Perform measurement of radiation levels for a radiation survey using an appropriate calibrated radiation survey meter required by subs. (1) (e) 1. and 2. A radiation survey meter shall also be used when there is potential for non-uniform exposure to personnel, such as may occur during machine maintenance or work in a radiation generating device target area.
DHS 157.87(3)(i)8. 8. During the first exposure for each set up of the device, measure the radiation levels around the perimeter of the controlled area. The perimeter shall be adjusted accordingly to meet the access control requirement for radiation areas or high radiation areas before subsequent exposures are made.
DHS 157.87(3)(i)9. 9. Conduct the survey around the perimeter for each new operating condition and adjust the perimeter accordingly. The area of operation shall be monitored periodically if radiation levels are variable.
DHS 157.87(3)(j) (j) Instruction and training. In addition to the requirements in sub. (1) (k), each individual permitted to operate or maintain an open-beam radiation generating device shall receive more specific and detailed instruction in and demonstrate competence in all of the following topics:
DHS 157.87(3)(j)1. 1. Sources and magnitude of common radiation exposure.
DHS 157.87(3)(j)2. 2. Units of radiation measurement.
DHS 157.87(3)(j)3. 3. Radiation protection concepts of time, distance, shielding, and ALARA.
DHS 157.87(3)(j)4. 4. Procedures and rights relating to a declared pregnancy.
DHS 157.87(3)(j)5. 5. Regulatory requirements and area postings.
DHS 157.87(3)(j)6. 6. Worker, embryo or fetus, and public dose limits.
DHS 157.87(3)(j)7. 7. Proper use of survey instruments and dosimetry.
DHS 157.87(3)(j)8. 8. The policies and procedures required by this subsection.
DHS 157.87(3)(k) (k) Personnel monitoring. In addition to the requirements of s. DHS 157.25, extremity dosimetry shall be provided and used by all of the following:
DHS 157.87(3)(k)1. 1. Personnel working with or routinely working near and having potential for exposure to, the primary beam of an open-beam radiation generating device.
DHS 157.87(3)(k)2. 2. Personnel maintaining radiation generating devices if the maintenance procedures require the presence of a primary radiation beam when any local component in the radiation generating device is disassembled or removed.
DHS 157.87(4) (4)Additional requirements for open-beam, hand-held radiation generating devices. In addition to the requirements in subs. (1) and (3) the following applies to open-beam, hand-held radiation generating devices:
DHS 157.87(4)(a) (a) Procedures. All registrants possessing open-beam, hand-held radiation generating device shall make their operating policies and procedures available for review by the department. Operating policies and procedures shall contain measures to ensure that all of the following occur:
DHS 157.87(4)(a)1. 1. Radiation protection is provided to meet public dose limits in s. DHS 157.23.
DHS 157.87(4)(a)2. 2. Radiation protection is provided equivalent to that afforded in sub. (3) (g).
DHS 157.87(4)(a)3. 3. The operator will not hold the sample during operation of the radiation generating device and the operator's hands will not approach the primary beam.
DHS 157.87(4)(a)4. 4. The operator will not aim the primary beam at themselves or at any individual during operation of the radiation generating device.
DHS 157.87(4)(a)5. 5. Operator radiation exposure is as low as reasonably achievable, for example, by use of ancillary equipment that will reduce exposure.
Loading...
Loading...
Published under s. 35.93, Stats. Updated on the first day of each month. Entire code is always current. The Register date on each page is the date the chapter was last published.