DHS 157.71(5)(e)(e) Whether the licensee notified the pregnant individual or mother, or the mother’s or child’s responsible relative or guardian, and if the licensee did not, whether such failure to notify was based on guidance from the referring physician.
DHS 157.71(6)(6)Records of instrument calibrations. A licensee shall maintain a record of instrument calibrations required by s. DHS 157.62 (1) for 3 years. The record shall include the model and serial number of the instrument, the date of the calibration, the results of the calibration and the name of the individual who performed the calibration.
DHS 157.71(7)(7)Records of radiation survey instrument calibrations. A licensee shall maintain a record of radiation survey instrument calibrations required by s. DHS 157.62 (2) for 3 years. The record shall include the date of the calibration, the results of the calibration, the name of the person who performed the calibration, and the model and serial number of the instrument.
DHS 157.71(8)(8)Records of dosages of unsealed radioactive material for medical use. A licensee shall maintain a record of dosage determinations required by s. DHS 157.62 (3) for 3 years. The record shall contain the radiopharmaceutical, patient’s or human research subject’s name or identification number if one has been assigned, the prescribed dosage, the determined dosage or a notation that the total activity is less than 1.1 MBq (30 microcuries), the date and time of the dosage determination and the name of the individual who determined the dosage.
DHS 157.71(9)(9)Records of possession of sealed sources and brachytherapy sources.
DHS 157.71(9)(a)(a) A licensee shall retain a record of leak tests required by s. DHS 157.62 (5) (b) for 3 years. The record shall contain the model number and serial number if one has been assigned of each source tested, the identity of each source radionuclide and its estimated activity, the results of the test, the date of the test and the name of the person who performed the test.
DHS 157.71(9)(b)(b) A licensee shall retain a record of the semi-annual physical inventory of sealed sources and brachytherapy sources required by s. DHS 157.62 (5) (g) for 3 years. The inventory record shall contain the model number of each source and serial number if one has been assigned, the identity of each source radionuclide and its nominal activity, the location of each source and the name of the person who performed the inventory.
DHS 157.71(10)(10)Records of surveys for ambient radiation exposure rate. A licensee shall retain a record of each survey required by s. DHS 157.62 (7) for 3 years. The record shall include the date of the survey, the results of the survey, the instrument used to make the survey and the name of the person who performed the survey.
DHS 157.71(11)(11)Records of the release of persons containing radioactive drugs or implants containing radioactive material.
DHS 157.71(11)(a)(a) A licensee shall retain a record of the basis for authorizing the release of a person for 3 years after the date of release if the total effective dose equivalent is calculated by any of the following methods:
DHS 157.71(11)(a)1.1. Using the retained radioactivity in the body rather than the radioactivity administered.
DHS 157.71(11)(a)2.2. Using an occupancy factor less than 0.25 at one meter to determine radiation exposure to persons physically near the patient.
DHS 157.71(11)(a)3.3. Using the biological or effective half-life of the radioactive material retained in the body.
DHS 157.71(11)(a)4.4. Considering the shielding by tissue to calculate the exposure to persons physically near the patient.
DHS 157.71(11)(b)(b) A licensee shall retain a record for 3 years after the date of release that the instructions required by s. DHS 157.62 (8) (b) were provided to a breast-feeding woman if the radiation dose to the infant or child from continued breast-feeding could result in a total effective dose equivalent exceeding 5 mSv (0.5 rem).
DHS 157.71(12)(12)Records of administrative and technical requirements that apply to the provision of mobile services.
DHS 157.71(12)(a)(a) A licensee shall retain a copy of the letter that permits the use of radioactive material at a client’s address of use, as required by s. DHS 157.62 (9) (a) 1., for 3 years after the last provision of service.
DHS 157.71(12)(b)(b) A licensee shall retain the record of each survey required by s. DHS 157.62 (9) (a) 4. for 3 years. The record shall include the date of the survey, the results of the survey, the instrument used to make the survey and the name of the person who performed the survey.
DHS 157.71(13)(13)Records of decay-in-storage. A licensee shall maintain a record of the disposal of licensed materials as required by s. DHS 157.62 (10) for 3 years. The record shall include the date of the disposal, the survey instrument used, the background dose rate, the dose rate measured at the surface of each waste container and the name of the person who performed the disposal.
DHS 157.71(14)(14)Records of contaminant concentration. A licensee shall maintain a record of the contaminant concentration tests required by s. DHS 157.63 (3) (b) for 3 years. The record shall include, for each measured elution or extract, all of the following:
DHS 157.71(14)(a)(a) The ratio of the measures expressed as kilobecquerel (microcurie) of molybdenum-99, strontium -82 or strontium-85 per megabecquerel of technetium-99m or rubidium-82 chloride injection.
DHS 157.71(14)(b)(b) The time and date of the measurement.
DHS 157.71(14)(c)(c) The name of the person who made the measurement.
DHS 157.71(15)(15)Records of instruction and training. A licensee shall maintain a record of instructions and training required by ss. DHS 157.64 (2), 157.65 (4) and 157.67 (4) for 3 years. The record shall include a list of the topics covered, the date of the instruction or training, the names of the attendees and the names of the persons who provided the instruction.
DHS 157.71(16)(16)Records of radiation surveys of patients and human research subjects. A licensee shall maintain a record of the surveys required by ss. DHS 157.65 (2) and 157.67 (2) for 3 years. Each record shall include the date and results of the survey, the survey instrument used and the name of the person who made the survey.
DHS 157.71(17)(17)Records of brachytherapy source inventory.
DHS 157.71(17)(a)(a) A licensee shall maintain a record of brachytherapy source accountability required by s. DHS 157.65 (3) for 3 years.
DHS 157.71(17)(b)(b) For temporary implants, the record shall include all of the following:
DHS 157.71(17)(b)1.1. The number and activity of sources removed from storage, the time and date they were removed from storage, the name of the person who removed them from storage and the location of use.